L63-RM1-PMR-SERVICE-ANALYZER
Wahyu Adi TrionoProblem: Sample line, filter, flange probe full material of coal, also filter balston
Action:
Clean up flange probe, sample line, filter probe, membran pump also disposable filter from dust coal process. Clean dryer gas and check filter humidity sensor replace filter with spare filter ( AGUS )
Status: done
Back to report list