Detail NR1 ELEC/INST RM/KL/CM
2024-08-15
L63-RM1
Wahyu Adi Triono
Problem: PMR service analyzer
Action:
PMR service analyzer. Clean up flange, sample line, filter probe, diaphragm pump. filter balston, dryer gas from dirty of coal. Check filter humidity sensor, replace filter with spare filter if found bad condition ( AGUS & ANDRI )

Status: done
Back to report list